Andersen Studies
Anprolene Door Left Open
This study documents the purge probe’s effectiveness in aerating the ls sterilization bag’s contents at the end of a sterilization cycle.
Within 45 minutes of the start of the 2-hour purge cycle, the EO samples decreased from approximately 40,000 ppm to 4 ppm – a reduction of 99.9%, and a measurement below the OSHA STEL (5 ppm). It can be concluded that by the end of the 2-hour purge cycle, the EO levels will be well below 4 ppm.